Model sizes/Description | 108 | 153 | 246 |
Stainless steel interior | Volume | approx. l | 108 | 153 | 246 |
Width | mm | 560 | 480 | 640 |
Height | mm | 480 | 640 | 640 |
Depth(less 25mm for fan) | mm | 400 | 500 | 600 |
Provision for grids or shelves half width/full width | number | - / 4 | - / 6 | 2x6 / 6 |
Textured stainless steel exterior | Width | mm | 710 | 630 | 790 |
Height (variable through adjustable feet) | mm | 778 | 938 | 938 |
Depth (withour door handle, depth of door handle 38mm) | mm | 550 | 650 | 750 |
Fully insulated, heated stainless steel door | | ○ |
Extra internal glass door | | ○ |
Ventilation | Uniform atmosphere and temperature distribution through enclosed non-turbulent ventilation system, fully covered by the sterilisation process | | ○ |
Temperature | Electronic microprocessor temperature controller with Pt100 and auto-diagnostic system | | ○ |
Temperature sensors Pt100 Class A in 4-wire circuit for uninterrupted operation on failure of one Pt1100 with warning indication | | double |
Temperature range(during sterilisation the temperature is fixed at +160°C -set value) | °C | from +20 to +50 (at least 8 above ambient) |
Temperature fluctuations with time (to DIN 12 880:2007-05) | K | ≤±0.1 |
Temperature variation in chamber at +37°C ( to DIN 12 880:2007-05) | K | ≤±0.3 |
Sterilisation | STERICard for automatic chamber sterilisation cycle 4 h at +160°C (not for sterilising the load) | | ○ |
CO2 | Digital electronic CO2 control with autozero, NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation | | ○ |
Setting accuracy | % CO2 | 0.1 |
Adjustment range | % CO2 | 0 to 10 |
Humidity | Capacitive humidity sensor(sterilisable) | | |
Standard water dishes | number | 1 | 2 |
Adjustable humidity limit control(88 - 97%)incl. digital indication and auto-diagnostic system with visual and acoustic fault indication(air supply via sterile filter)ensures rapid reaching of set humidity and short recovery times while avoiding condensate formation | | ○ |
Monitor | Microprocessor temperature monitor acting as overtemperature protection(protection class 3.1),with Pt100 incorporating fault diagnostics with visual and audible alarm | | ○ |
Digital over- and undertemperature monitor | | ○ |
Temperature monitoring band automatically linked to the setpoint(ASF) | | ○ |
Relay for reliable heating cut-off in case of fault | | ○ |
Mechanical temperature limiter(TB) | | ○ |
Audible alarm:Over- and undertemperature, over-CO2 and empty gas cylinder, open door, underhumidity and empty water tank(with optional Humidity module) | | ○ |
Timer functions | Real-time/weekly programmer with group function(e.g. Monday - Friday) | | ○ |
Setup | Calibration(no separate PC required), Temperature:3-point calibration on controller, CO2:3-point calibration at 5%, 7% and 10%, Auto-zero-function of NDIR CO2-sensor after every sterilisation and cyclically every 24h, Humidity: 2-point calibration at 20% and 90% | | ○ |
Setting of language for dialogue and display DE/EN/ES/FR/IT | | ○ |
Further data | Electrical load at 230V, 60/60Hz | approx. W | 1000 | | |
Standard accessories | Perforated stainless steel shelves(full width) | number | 2 | 3 |
Stainless steel wate dishes, 40mm high | number | 1(full width) | 2 (half width) |
Works calibration certificate(measuring point chamber centre at +37°C) | | ○ |