Range |
0-5 up to 20,000 mg/L TOC |
Accuracy |
±2% of reading, non-diluted ranges at 25℃ (77˚F) ±4% of reading, diluted
ranges at 25℃ (77˚F) |
Minimum Detection Limit |
≤0.015 mg/L for range 0-5 mg/L at 25℃ (77˚F) |
Response Time |
T90 ≤ 8 minutes (includes TIC sparging) |
Inlet Pressure |
0.15-6 bar (2-87 psig) |
Flow Rate |
20-200 mL per minute |
Sample Temperature Range |
2 to 70℃ (36 to 158˚F) |
Operating Temperature Range |
5 to 40˚C (41 to 104˚F) |
Recorder Outputs |
Two 4-20 mA analog outputs selectable for sample concentration, analy system
warning or auto range indication |
Alarms |
Five alarms selectable for sample concentration alarm, analyzer system warning
or analyzer system shutdown alarm Each is equipped with an SPDT relay with
contacts rated for 3A resistive load at 250 Vac or 0.5A at 30V |
serial Communication |
One multi-function RS232 or RS485 serial port (ModBUS, CSV) |
Power |
115/230 Vac 50/60 Hz (switch selectable), 300 VA maximum |
Sample Inlet/Outlet Connection |
1/4-inch OD tube, compression fitting |
Drain Connection |
1 1/2-inch OD standard drain pipe |
Carrier Gas |
1/8-inch OD tube, compression fitting Clean, Co2 free air or Nitrogen
at 2.8-6.2 bar(40-90 psig) |
Compliance/Certification |
UL/CSA, CE approved Standard Methods 5310 C, EPA 415.1 |